Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker 1288 Bedienungsanleitung Seite 9

Vorschau ausblenden Andere Handbücher für 1288:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
5.
Soak the device in the detergent for at least
15 minutes.
Cleaning: Manual
1.
Brush
• Thoroughly brush the exterior of the device
with a soft-bristled brush, focusing on any
mated or rough surfaces.
• Inject any lumen or mated surface a
minimum of five times with at least
50mL of the detergent.
• Brush any lumens a minimum of five
times from each end, using an appropriate
bottle brush.
2.
Rinse
• Rinse the device with treated water at
ambient temperature to remove all detergent
residue. Flush any lumens or mated surfaces
a minimum of five times. Once all detergent
residues have been removed, continue to rinse
for a minimum of 30 seconds.
• Drain excess water from the device and dry it
using a clean cloth or pressurized air.
• Visually inspect the device for cleanliness,
paying close attention to hard-to-reach areas.
If visible soil remains, repeat steps 1 and 2.
3.
Soak
• Prepare a non-enzymatic detergent, according
to the manufacturer's recommendations of
0.25 ounces/gallon tap water at 35 – 40°C.
• Fully immerse the device and inject any
lumens and mated surfaces with at least
50mL of the detergent.
• Soak the device for a minimum of 15 minutes.
4.
Brush
• Thoroughly brush the exterior of the device
using a soft-bristled brush.
• Inject the prepared detergent into any
cannulas, lumens, or mated surfaces
a minimum of five times.
• Brush any lumens a minimum of five
times from each end, using an appropriate
bottle brush.
• Actuate the device, brushing around any
movable parts in all extreme positions.
5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis