Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker 1288 Bedienungsanleitung Seite 10

Vorschau ausblenden Andere Handbücher für 1288:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
5.
Rinse
• Thoroughly rinse the device with treated
water until all detergent residue is removed.
Flush any lumens or crevices five times. After
the detergent residue is removed, continue
rinsing for a minimum of 30 seconds.
• Drain the excess water from the device and
dry it with a clean cloth or pressurized air.
Cleaning: Automated
1.
Brush
• Brush both ends of any lumens a minimum of
five times, using an appropriate bottle brush.
2.
Rinse
• Rinse the device with treated water at ambient
temperature until there is no visible detergent
residue. Continue to rinse for a minimum of
30 seconds after all detergent residue has been
removed.
• Place the device in the washer on an incline to
facilitate drainage.
3.
Automated wash
• Program the washer using the following
parameters:
Pre Wash
Enzyme
Wash
Wash I
Rinse I
Dry
Phase
• If necessary, use pressurized air to aid in
drying. Visually inspect each device for
cleanliness.
6
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Cold
Hot
Set point
66°C
(151°F)
Hot
115°C
(239°F)
N/A
Enzymatic
Detergent
(1 oz/gallon)
Regular
Detergent
(.25 oz/gallon)
N/A
N/A