Disinfection (optional)
1.
Disinfect the device by submerging it in
a disinfection agent, such as
• ≥ 2.4% glutaraldehyde
or
• ≥ 0.55% ortho-phthalaldehyde.
Follow the disinfection agent
manufacturer's recommended
concentrations, temperatures,
and exposure times.
2.
Thoroughly rinse and flush all parts and
lumens with running, demineralized water
to remove the disinfectant.
3.
Dry all parts with a lint-free towel
immediately after rinsing.
Drying
• For automated drying, use the drying cycle
provided with the washer/disinfector.
• For manual drying, use a lint-free cloth.
• Dry any lumens with compressed air.
Maintenance, Inspection, and Testing
• Inspect the device on a continual basis.
If a problem is observed or suspected,
the device should be returned for repair.
• Inspect all components for cleanliness. If fluid
or tissue buildup is present, repeat the above
cleaning and disinfection procedures.
• Inspect the device for cuts and breaks. Return
any damaged device to Stryker for service.
Packaging
N/A
Sterilization
After performing the cleaning instructions
specified above, perform one of the following
sterilization cycles.