Micro Cart
REF 351-600-000
PURPOSE:
The purpose of the Micro Cart is to house
and transport medical equipment.
Tools Needed:
Transport and Storage Conditions
Temperature: -40 – 70°C (-40 – 158°F)
Relative Humidity: 10 – 85%
MEDICAL CART
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL60601-1, IEC60601-1,
CAN/CSA C22.2 No. 601.1
39A3
EN-1
, Inc.
1/8" Hex Wrench
IMPORTANT INFORMATION: File in your
CLEANING AND MAINTENANCE:
CLEANING:
Wipe surfaces with soap and water or disinfectant.
Dry with non-abrasive cloth or paper towel.
To ensure ease of use, keep casters free of dirt, oil, and
debris.

MAINTENANCE:

Periodically check all hardware and re-tighten fasteners
as required.
Keep debris from wheel areas.
Check cart for loose, damaged or worn components and have
a qualified technician replace as required.

SAFETY PRECAUTIONS:

Distribute load evenly in drawer and on shelves.
DO NOT overload drawer or shelves, refer to weight load
limits.
Shelf / drawer load limit: 25 pounds each.
Total cart capacity fully loaded: 75 pounds.
Always bottom load for optimal stability.
I.V. Pole load limit of 6000 ml of saline.
Lock wheels when not in transport. Release brakes when
transporting.
Use caution when rolling the cart over uneven surfaces such
as floor mats or door and elevator thresholds, as contents may
shift or tip.
Never leave cart unattended on incline surfaces.
Do not attempt any modifications or repairs without proper
training.
IV Hanger Assembly
Tighten set screw from IV
hanger assembly onto IV pole
WARNING:
DO NOT   transport cart with IV Pole extended or loaded. This
could result in a unbalanced load, causing cart to tip, and result
in injury or product damage.
Arthroscopy Pump Calibration: Saline Bag IV Pole must be 1 m
(3 ft.) above pump. Failure to do so may cause inaccurate pump
calibration.
www.stryker.com
maintenance records
Insert IV Pole
Insert IV pole into
hole in the top of the cart
1000-400-939 E