Important
This booklet is designed to assist in using this product. It is not a reference to surgical
techniques.
Device description
The Ultravision
Euro Patient Return Adaptor is intended for use with the Ultravision
TM
and 'Euro style' patient return electrodes. The Ultravision
return electrode in order to function. The Euro Patient Return Adaptor allows the Ultravision
Generator to share the same patient return electrode as a monopolar electrosurgical instrument
that is fitted with a 'Euro style' patient return electrode socket.
The Ultravision
Euro Patient Return Adaptor is only necessary when using the Ultravision
TM
System with a monopolar instrument. When using a bipolar or ultrasonic instrument, the patient
return electrode with an 'international style' plug may be connected directly to the return socket
in the fascia of the Ultravision Generator. If such an electrode is unavailable the Euro Patient
Return Adaptor can be used as an adaptor to allow a 'Euro style' patient return electrode to be
connected to the return socket of the Ultravision™ Generator. Where multiple instrument types
may be required during the procedure, if one of these is a monopolar instrument then the Patient
Return Adaptor should be used.
The Euro Patient Return Adaptor is supplied non-sterile and does not require sterilisation.
The Euro Patient Return Adaptor should be replaced after 500 mating cycles.
Indications
The Ultravision System is indicated for the clearance of smoke and other particulate matter that
is created during laparoscopic surgery by monopolar, bipolar and ultrasonic surgical instruments.

Contraindications

The Ultravision
system is an electrical energy-based system that must not be used on patients
TM
fitted with a pacemaker or implantable cardioverter defibrillator (ICD).
Warnings and Cautions
• The Ultravision
System should only be used by appropriately trained medical personnel.
TM
• Always disconnect the Euro Patient Return Adaptor from the electrosurgical generator and
Ultravision
Generator before cleaning.
TM
• Do not apply alcohol, caustic, corrosive, or abrasive cleaning or disinfectant compounds,
solvents, or other materials that could cause damage. See Cleaning section below.
• Do not attempt to sterilise the Euro Patient Return Adaptor.
• Keep the Euro Patient Return Adaptor away from liquids.
How to Use the Ultravision™ Euro Patient Return Adaptor – Before surgery
Consult the User Manual for detailed instructions on how to use the Ultravision System.
1.
Connect the patient return electrode to the patient using standard surgical practice.
2.
Connect the patient return electrode to the Euro Patient Return Adaptor (Figure 1).
3.
Connect the patient return adaptor to the electrosurgical generator (Figure 2).
4.
Switch on the electrosurgical generator and ensure that the patient return path is
Generator requires a patient
TM
2
System