Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo InTap Benutzerhandbuch Seite 49

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
Scope of delivery
• Power supply 30 W: AC/DC adapter with AC adapter set EU, US, AU, UK
• USB stick, plugged-in the USB connection
Supply voltage (Power)
You have the following possibilities to power the device:
• either via the internal battery
• or via the supplied AC/DC adapter connected to 100 to 240 V AC / 50 to 60 Hz.
Hoses "Medium In" and "Medium Out"
You can connect the following hoses to "Medium In" and "Medium Out":
InTap portable oDO analyzer
Description
With metric connectors
With imperial inch /
US connectors
Adjusting medium flow rate
The flow rate is adjusted with the rotary knob. The flow rate depends on the medium pressure
The flow rate increases with increasing medium pressure.
1. Turn rotary knob clockwise to the end stop.
2. Turn rotary knob anti-clockwise to adjust the flow rate. The rotary knob has a click-stop
position every 15°. E.g. for a flow rate of approximately 200 ml/min at 2 bar you have
to turn the rotary knob to the 30° position. That means you hear 2 clicks.
For the diagram "Adjusting the flow rate" refer to the Reference Manual.
InTap Portable oDO Analyzer • 30 461 735 A
Order number
30 425 550
30 457 912
Hoses
Description
6 mm
1/4" NPT
Order number
30 422 575
30 432 818
49
en

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis