OPERATION

In case of emergency, disconnect the suction unit from the mains (pull out the mains
plug).
Parts of the suction unit can have hot surface areas. There is the danger of scorching if
contact is made.
The suction unit operates after receiving a signal from the dental chair. The cooling fan
and suction pump always work together.

12. SWITCHING SUCTION UNIT ON

When the suction unit is switched on at the plug, power is provided to the switching block, and puts the
aspirator - on stand-by mode - waiting for a signal from the dental chair. When a signal is received, the
suction unit operates and creates a vacuum in the inlet path until a signal interrupts its operation, and puts
the suction unit back - on stand-by mode.

MAINTENANCE

13. MAINTENANCE SCHEDULE

Required maintenance
• Separator - working correctly
( suction path free of liquids – dry )
• Full service of suction system
14. MAINTENANCE
Repair work beyond normal maintenance can be performed only by qualified personnel or
the manufacturer's representative.
Use only spareparts and accessories approved by the manufacturer.
Prior to any maintenance or repair work, switch off the compressor and disconnect it from
the mains (pull out the mains plug).
Prior to any maintenance or repairs, the soundproof box must be removed.
The suction unit is designed and manufactured to keep maintenance to a minimum.

15. STORAGE

If the appliance is not to be used for a prolonged period of time, the manufacturer recommends operating the
entire aspirating system for approximately 10 minutes, with an unrestricted flow of air (without cannules).
Then disconnect the appliance from the mains. Cover the inlet and outlet ports with the plastic plugs that
were supplied with the unit
NP-DO2.1_4_01-2011
See document
Frequency
Documentation
of dental unit
Service
documentation
- 18 -
DO 2.1-10
Performed by
monthly
yearly
qualified
engineer
DO 2.1
user
09/2010