To reduce the risk of injury from possible end user exposure associated with a) biological contamination and b)
degradation products released from the purifi cation media; and
To reduce the risk of product loss associated with a) chemical release from purifi cation media, and b)
nullifi cation of Supplier's USP<88> Class VI compliance of fi lter capsule components:
• During pre-use autoclave sterilization do not exceed a temperature of 121°C or a duration longer than 30 minutes for
numbers BV8R Luer (EMP301AEX020R) and BV8R Sanitary (EMP303AEX020R).
• Do not use with or expose this product to process liquids until capsule temperature goes <40°C (104°F).
• Only use NaOH as sanitizing chemical and do not exceed 1M NaOH concentration or 1-hour duration for sanitization.
• Do not NaOH sanitize 3M Emphaze AEX Hybrid Purifi er and 3M Zeta Plus Encapsulated capsules together in series.
• Use only 70% ethyl and/or isopropyl alcohol to wipe the exterior of the capsules.
• Do not use with or expose this product to pressurized steam or sanitizing chemicals other than NaOH.
To reduce the risks associated with chemical exposure, impact, crush, tipping hazards, property damage from
product loss, bypass or cross contamination due to pressure barrier failure, connector failure, purifi cation media
or capsule failure:
• Use the installation procedure specifi ed in Operation Instructions.
• Wear Personal Protective Equipment (PPE) to prevent chemical splash.
• Hardware Qualifi cation testing is recommended prior to use.
• The capsule must be subjected to a preconditioning fl ush prior to use. Use the procedure specifi ed in Operation Instructions.
• After pre-use NaOH sanitization, fl ush the system with appropriate aqueous buff er until the pH range of 5 to 9 has
been attained and monitor the pH of the capsule effl uent to determine when the baseline should be reached
(e.g., ± 0.5 pH units from the pH of the buff er used to fl ush the media).
• Capsules are designed for single use only. Do not reuse.
• Do not modify or alter the capsules.
• Do not put this product into service if any damage is occurred to the capsule or any of the connecting parts.
• Do not install where fl uid pressure exceeds 2.8 bar (40 psig).
• Do not use in organic solvent based processing.
WARNING
WARNING
- 3 -