Product Specifi cations
Miscellaneous
Connector
Recommended Use Flow Rate
Maximum Flow Rate
6
Storage Conditions
7
Shelf-Life
8
1
Do not use this product for continuous service with compressed gasses. Use of compressed gas is permissible for post-use
integrity testing and blow-down purposes only. Limit post-use integrity test and blow-down gas pressures to less than 1.4 bar
(20psig) and temperature less than 25°C for no longer than 30 minutes.
2
Capsule Fill Volume is defi ned as the volume of liquid that is required to fi ll the capsule.
3
Post Blow down Hold up Volume is defi ned as the volume of the residual liquid after gas (air or nitrogen) blow down.
4
The Required Preconditioning Flush is a saline solution, having a conductivity in the same range as that of 25 to 150 mM NaCl (typical
3-16 mS/cm). DO NOT use water to fl ush the capsule.
5
Review Regulatory Support File for recommended alternate buff er selection.
6
Do not allow the liquid diff erential pressure across the capsule to exceed 2.4 bar (35 psid). Maximum fl ow rate may be limited by
system diff erential pressure.
7
Do not allow product to be exposed to freezing or extreme heat conditions.
8
USP <88> Class VI compliance and high adsorption capacity for up to 2 years from date of manufacture if stored as indicated. Product
labeling includes the expiration date.
Do not steam-in-place sterilize the capsules.
Preconditioning Flush Procedure
The Laboratory Capsule must be fl ushed prior to use to remove the glycerin, that is present as a media stabilizer. Flush the capsule
with 54 L/m
2
(15.3 mL/bed volume) of a saline solution or buff er having a conductivity in the same range as that of 25 to 150 mM
NaCl (3 - 16 mS/cm) at a fl ow rate no higher than 210 L/m
fl ow direction only (see Figure 2). Intermittently bleed air off of the upstream side of the capsule using the Luer Capped Air Vent on
the inlet side of the capsule.
BV8R Luer
EMP301AEX020R
Luer
8 mL/min
24 mL/min
Controlled indoor temperature: 0 - 30°C (32 - 86°F) in original packaging
2 years from date of manufacture @ 30°C maximum storage temperature
WARNING
2
/h (LMH). DO NOT use water as the fl ush solution. Flush in the forward
- 8 -
BV8R Sanitary
EMP303AEX020R
¾" Sanitary Connectors