Herunterladen Inhalt Inhalt Diese Seite drucken

Invacare SoftCloud Ace Gebrauchsanweisung Seite 32

Wechseldrucksysteme
Vorschau ausblenden Andere Handbücher für SoftCloud Ace:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 73
SoftCloud® Ace, SoftCloud® Pro
Electromagnetic Immunity
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/ burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
32
IEC 60601 test level
±8 kV contact
±15 kV air
±2 kV for power supply lines
± 1 kV line(s) to line(s)
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Compliance level
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV differential mode
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the mattress system requires
continued operation during power mains
interruptions, it is recommended that
the mattress system be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
1630732-B