Operating conditions
Medium conditions
● Process temperature
Measuring cell with silicone oil
filling
Mechanical design
Material
● Wetted parts materials
Seal diaphragm
Process connection and locking
screw
O-ring
● Non-wetted parts materials
Electronics housing
Process flange screws
Mounting bracket
Display and user interface
Terminal voltage at transmitter

Certificates and approvals

Certificates and approvals
Classification according to Pressure
Equipment Directive (PED 97/23/EC)
PN 160 (MWP 2320 psi)
Explosion protection
● Explosion protection for Europe (to
ATEX)
● Intrinsic safety "i"
Marking
Permitted ambient temperature
Connection
Effective inner inductance
Effective inner capacitance
● Flameproof enclosure encapsulation
"d"
Marking
Permitted ambient temperature
16
-40 ... +125 °C (-40 ... 257 °F)
(Note that there is no medium flow through the pressure caps during normal
operation at a temperature > 100 °C.)
Stainless steel, mat. no. 1.4404/316L
PN 160: Stainless steel, mat. no. 1.4404/316L
FPM (Viton), optionally NBR
● Non-copper aluminum die casting AC-AlSi12 (Fe) or AC-AlSi 10 Mg (Fe)
as per DIN EN 1706
● Polyurethane-based paint, optional with epoxy-based primer
● Nameplate made of stainless steel, mat. no. 1.4404/316L
Stainless steel, mat. no. 1.4404/316L
Steel, mat. no. 1.0330 or stainless steel, mat. no. 1.4301
● 10.6 V ... 44 V DC
● In the case of intrinsically safe operation 10.6 V ...
● For gases of Fluid Group 1 and liquids of Fluid Group 1; meets
requirements of Article 3 Para. 3 (good engineering practice)
PTB 09 ATEX 2004 X
II 1/2 G Ex ia/ib IIC T4
-40 ... +85 °C (-40 ... +185 °F) temperature class T4
To a certified intrinsically safe circuit with the max. values:
U
= 30 V, I
= 100 mA,
i
i
P
= 750 mW, R
= 300 Ω
i
i
L
= 400 μH
i
C
= 6 nF
i
BVS 09 ATEX E 027
II 1/2 G Ex d IIC T4/T6
-40 ... +85°C (-40 ... +185°F) temperature class T4
-40 ... +60°C (-40 ... +140°F) temperature class T6
30 V DC