Herunterladen Diese Seite drucken

Stryker 1488 Handbuch Seite 37

Hd video camera volume l of ll
Vorschau ausblenden Andere Handbücher für 1488:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 99
Ethylene Oxide (EO)
1.
Clean and prepare the camera head and cable as recommended in this
user guide. Ensure the soaking cap is installed.
2.
If using a sterilization tray (optional), follow any additional instructions
provided with the tray. Use only trays that are compatible with EO
3.
Double wrap the camera head and cable (or tray) prior to sterilization.
4.
Sterilize the camera head and cable using the parameters below.
5.
Allow the camera head, cable, coupler, and scope to completely dry
before reassembly. Any moisture on the threads will cause the camera
and coupler windows to fog during use.
Preconditioning parameters
Temperature
Chamber Humidity
Vacuum Set Points
Time
Exposure
Concentration (100% EO)
Temperature
Time
Chamber Humidity
Aeration parameters
Aeration Time
Temperature
55 °C (131 °F)
70% RH
1.3 psia
30 minutes
725 mg/L
55 ± 2 °C (131 ± 5 °F)
1 hour
70% RH (50–80%) ± 5%
12 hours
35–54 °C (95–129 °F)
.
8
EN-33