4.
Select the same operating frequency for the receiver.
• Press the set button on the receiver two times.
• Press the up or down arrows to select the same frequency selected
on the transmitter.
• Hold the set button until the LCD shows "Stored. "
• Press the down arrow until "Quit" shows on the LCD.
• Press the set button until the status bar shows on the LCD.
5.
Clip the transmitter to an article of clothing away from the hands and
operating site.

Removing the Cartridge from SIDNE

If the upgrade cartridge needs to be removed, do the following:
1.
Power off SIDNE.
2.
Disconnect the DVI cable from the voice box cartridge.
3.
Turn the thumbscrews counterclockwise.
4.
Pull the cartridge out from the back of SIDNE.
5.
Reattach the metal plate to cover up the exposed slot by turning the
thumbscrews clockwise until the plate is locked into place.