Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 13

Operation

Before operating the VWR MiniStar centrifuge, be sure that the power switch on the back of the
unit is in the"on" position. To begin a run, simply close the lid of the centrifuge. No other controls
are required. To stop rotation, press and release the tab on the front of the lid. This will release
the lid lock, allowing the lid to self open and the rotor to come to a gradual stop. Caution: Do not
attempt to open the lid or remove samples until the unit has come to a complete stop. Once the lid
has fully opened and the rotor has stopped, samples may be removed from the rotor.
Safe operation of the centrifuge requires that the rotor be loaded in a balanced fashion. Failure to
load the rotor properly presents a hazardous condition and may damage the centrifuge. Individual
tubes, strip tubes and/or adapters must always be loaded symmetrically to ensure proper balance.
Never run the centrifuge with only one strip or tube in place.
The VWR MiniStar centrifuge is supplied with two interchangeable quick release rotors. No tools
are required to exchange the rotors or attach them to the motor shaft. The rotors can be easily
exchanged with one hand by way of the quick release system. Removal of the rotors is done by
placing your thumb on top of the motor shaft and lifting up with two fingers under the rotor. Place
the other rotor into position on the motor shaft and snap in.
(8-place tube rotor and 4-place strip rotor)
3 | VWR MiniStar Centrifuge

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ministar blueline

Inhaltsverzeichnis