Herunterladen Inhalt Inhalt Diese Seite drucken

Control Panel Description; Operation - VWR VAPOUR-Line eco 50 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24

Control panel description

The autoclave is equipped with specially designed microprocessor that controls the entire instrument and the
sterilisation cycles.
Pos. 1 - Display
Pos. 2 - Start button
Pos. 3 - Stop button
Pos. 4 - Menu button
Pos. 5 -  button
Pos. 6 -  button
Pos. 7 - Enter button
Pos. 8 - LED

Operation

1. The autoclave should be placed in a room accessible by the operator only
2. Check that the heating element inside the vessel is fully covered by water
3. Check that the water tap (Pos. 8 page 7) is closed
4. Connect a suitable container to the hose connection of the condensate discharge port (Pos. 8 page 7)
Multipurpose, it shows all the system messages and sterilisation
cycle parameters such as temperature and time
Starts the sterilisation cycle shown on the display
Stops the sterilisation cycle and resets all the alarms
WARNING! STOP button is an emergency button. Pushing it will
abort the cycle and will open the valves. We don't recommend
you push it unless a problem occurs during the cycle.
The same thing will happen if the instrument is turned off by the
main switch.
Enters the instrument setting menu
Scrolls up or increases the displayed value
Scrolls down or decreases the displayed value
Confirms and stores the displayed value
Reports the status of the autoclave
PRINTER ON: Lit when the printer is enabled
END CYCLE: Lit at the end of a sterilisation cycle
ALARM: lit when an alarm occurs
8

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis