Press PUMP CURRENT push-button and the
display shows:
E X T E R N A L
I N P U T = X X . X
This screen will only be displayed if YES was selected in
EXTERNAL INPUT.
where: XX.X = is the voltage rating measured on
connector P34 0 to 10 Vdc.
Press the CYCLE NUMBER twice and the display
shows:
X X X X C Y C L E
P U M P
L I F E
where:
CYCLE = are the cycles performed (range 0
to 9999)
m = is the elapsed time related to the cycle
number
displayed
minutes)
PUMP LIFE = is the total operation time of
the pump (range 0 to 99999 hours).
V
X X X X X m
X X X X X h
(range
0
to
99999

Front / Remote / Serial Selection

Press CYCLE NUMBER and PUMP CURRENT
pushbuttons together for at least 2 seconds and the
processor enters in a routine where it is possible to
program the controller.
In this routine, the CYCLE push-button is used for
choosing/changing the value or condition;
the PUMP CURRENT push-button is used to enter and
confirm the value.
At any time it is possible to exit this routine by pressing
the CYCLE and PUMP CURRENT pushbuttons at the
same time for at least 2 seconds.
The display shows:
C O N F I G U R A T I O N
S O F T W A R E
Q E X X X X X X X X X
and then:
S O F T
S E L E C T I O N :
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to
rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected
and the rump-up of the pump will be done within 25
minutes.
The controller is factory set to YES.
The Soft Start mode may be deselected/selected only
when the pump is stopped.
63
TECHNICAL INFORMATION
V E R S I O N
S T A R T
M O D E
X X X
NOTE
87-900-855-01 (B)