Herunterladen Inhalt Inhalt Diese Seite drucken

Optional Serial Port; Serial Port Installation; Serial Communication Port J13 And J14; Rs 232-422 Communication Descriptions - Varian Turbo-V 6000 969-9491 Bedienungshandbuch

Inhaltsverzeichnis

Werbung

OPTIONAL SERIAL PORT

Serial Port Installation

An optional R S 232/422/485 kit is available for both
models. To install it, proceed as follows:
Switch off the power.
Unscrew the cover screws and remove the cover.
Unscrew screw A and remove board B.
Attach the interface board connector to connector
P6 on the board B and then secure it in place using
screw C (3x6), which is not provided.
Reinstall board B and lock it to the controller
through the screw.
Insert the flat cable through the rear panel and plug
it into the socket P1 located on the Interface board.
Restore the main board into its original position.
Install and tighten the connector screws and turrets.
Replace the cover.
Serial port connector installation

Serial Communication Port J13 and J14

Communication serial port connections and mini- mum
connection configuration are shown in the following
figures. The communication port mating connector is
supplied with the RS 232 PCB (AMP/Cannon or
equivalent 9-pin "D" type male connector). The external
cable (not supplied) between the host computer and
the controller does not require crossed wires so that
signals are connected correctly.
For example, the Transmit data signal from controller
(pin 2) must be connected to the host computer's
Receive data line (pin 2) and vice versa. Consult the
host computer's instruction manual for its serial port
connections.
Varian
cannot
guarantee
regulations for radiated emissions unless all external
wiring is shielded, with the shield being terminated to
the metal shroud on the O-subconnector. The cable
should be secured to the connector with screws.

RS 232-422 Communication Descriptions

Communication RS 232 serial port connections
Communication RS 422 serial port connections
Transmission Channel Characteristics
levels:
baud rate:
character length:
parity:
stop bit:
protocoll:

Message Structure

(request and answer have the same format)
The master system (PC) starts every session sending
the following message to the slave units connected:
<STX> / <ADR> + <WINDOW> + <COMMAND> +
<DATA> + <ETX> + <CRC>
56
TECHNICAL INFORMATION
NOTE
compliance
RS232/RS422
9600/4800/2400/1200/600
programmable
8 bits
none
1 bit
master (PC) / slave (converter)
87-900-855-01 (B)