INSTALLATION
!
The Turbo-V controller is equipped with a 3-wire power
cord and plug (internationally approved) for user safety.
Use this power cord and plug in conjunction with a
properly grounded power socket to avoid electrical
shock.
High voltage developed in the controller can cause
severe injury or death. Before servicing the unit,
disconnect the input power cable.
The Turbo-V controller can be used as a bench unit or
a rack module, but it must be positioned so that free air
can flow through the holes.
Do not install or use the controller in an environment
exposed to atmospheric agents (rain, snow, ice), dust,
aggressive gases, or in explosive environments or
those with a high fire risk.
During
operation,
the
conditions must be respected:
temperature: from O °C to +40 °C;
relative humidity: 0 - 95%
(without condensation).
To connect the controller to the pump use the specific
cable supplied with the controller.
See the appendix "Technical Information" for detailed
Information about the above mentioned and the other
connections, and about the options installation.
USE
This paragraph describes the fundamental operating
procedures.
Detailed
procedures that envolve optional connection or option
are supplied in the paragraph "USE" of the appendix
"Technical Information".
Make all vacuum manifold and electrical connections
and refer to Turbo-V pump instruction manual before
operating the Turbo-V controller.
!
To avoid injury to personnel and damage to the
equipment, if the pump is laying on a table make sure it
is steady .
Never operate the Turbo-V pump if the pump inlet is
not connected to the system or blanked off.
WARNING!
NOTE
following
environmental
information
and
operating
WARNING!
The input signal J1connector should be left in position
including the shipping links if no external connections
are made. The forepump and Turbo-V pump can be
switched on at the same time.

Controller controls, indicators and connectors

The following paragraph illustrates the Controller
control panel and interconnection panel.
More details are contained in the appendix "Technical
Information".
1. Keyboard
push-button
selection. It is active only when the front panel
operation has been selected. Pressed once, the pump
runs at about 2/3 of the nominal speed. To unselect
the mode, press the push-button again.
2. Keyboard push-button for START, STOP, RESET
mode selection. It is active only when the front panel
operation has been selected. By pressing once the
starting phase begins; if pressed again it stops the
pump. If the pump has been stopped automatically by
a fault, this push-button must be pressed once to reset
the controller and a second time to restart the pump.
3. Keyboard push-button to recall on the display the
cycle number, cycle time and pumplife.
4. Keyboard push-button to recall on the display the
pump current, pump power, rotational speed and lower
bearing temperature, upper bearing temperature,
motor temperature, excitacion frequency. It is always
active regardless of the operating mode selected.
Push-buttons 3 and 4, if pressed together for at least
2 seconds put the controller in a routine where it is
possible to program some operation parameters.
5. LCD back-lighted alphanumeric display: dot matrix 2
lines x 24 characters.
Controller 969-9491,
969-9591 front panel
46
INSTRUCTIONS FOR USE
NOTE
for
LOW
SPEED
mode
87-900-855-01 (B)