Connection P1
Logic Input Interconnections
P1 input connector
All the logic input to the controller must be connected at
J-1 mating connector. With the provided J1 mating
connector (shipped with pin 3 and pin 8 shorted) make
the connections with AWG 24, (0.24 mm ) or smaller
wire to the pins indicated in the figure to obtain the
desired capability. The following table describes the
signals available on the connector.
PIN
DESCRIPTION
1-6
Remote START/STOP optically isolated from
the internal circuit, requires a permanently
closed contact (relay contact transistor, etc. ).
When the contact closes. the turbopump starts,
and when the contact opens. the turbopump is
stopped. With the remote mode operation
selected,
the
inoperative.
2-7
Remote LOW SPEED optically isolated from
the internal circuit, requires a permanently
closed contact (relay contact, transistor. etc.).
When the contact closes, the turbopump runs
at low speed and when the contact opens the
turbopump reverts to high speed mode. With
the remote mode operation selected, the front
panel push-button is inoperative.
3-8
INTERLOCK
optically
internal circuit, this signal can be used to delay
the starting of the turbopump.
permanent closed contact before starting the
turbopump.
4-9
SYSTEM OVERRIDE optically isolated from the
internal circuit, this signal is used to stop the
pump in emergency condition, requires a
closed contact.
When the contact is closed, the turbopump
and the interconnected devices are stopped.
front
panel
push-button
isolated
from
requires a
Pin 3-8 must be shorted to allow the Turbo-V 6000
pump to start if no interlock contact is connected. if,
after starting the pump, the interlock contact opens, it
has no effect on the operation and the pump continues
to turn.
The following figure shows a typical contact logic input
connection and the related simplified circuit of the
controller.
is
the
53
TECHNICAL INFORMATION
NOTE
Typical logic input connection