VisionAire™
5.0
Troubleshooting
If your concentrator fails to operate properly, consult your Equipment Provider,
and refer to the troubleshooting chart on the following pages for probable causes
and solutions.
Do not attempt any maintenance other than the possible solutions
listed below.
If the unit has not been used for an extended time period, it needs
to operate for several minutes before power failure alarm can
become activated.
Problem
Unit does not
operate. Power
failure condition
causes an alarm to
sound.
®
AirSep
Corporation
MN137-1 rev. C - 02/13
Probable Cause
The power cord is not
connected to the
electrical outlet.
No power at the
electrical outlet.
Circuit breaker on
concentrator unit is
activated.
Solution
Check power cord at the
electrical outlet for a proper
connection.
Check power source, wall
switch, in-house fuse or circuit
breaker.
Press (do not hold in) the circuit
breaker reset button in the front
of the unit.
If the circuit breaker trips again
or the alarm continues to sound
after the unit is turned on,
contact your Equipment
Provider.
EN
EN-21