Calibrate BP
Setup Menu
To callibrate BP-cuff
Shake Sensor
Display Backlight
Inflate cuff to
Display Contrast
100 mmHg
Autopower off
Then press "Calibrate"
Com. Channel
Calibrate BP
Language
To ensure correct measurements of simulated BP, the
sphygmomanometer used with the cuff needs to be calibrated
to the pressure sensor in the simulator.
This is done by selecting the Calibrate BP function, inflating
the cuff to 100mm Hg and pressing the Calibrate quick execution
button as the pressure is held at exactly 100mm Hg.
Language
Setup Menu
1 Deutsch
Shake Sensor
Display Backlight
2 English
Display Contrast
3 Espanol
Autopower off
4 Francais
Com. Channel
5 Italiano
Calibrate BP
Language
6 Nederlands
7 Norsk
Language selection affects the texts on the remote control display.
Selection of a new language will also affect the "Yes" and "No"
vocal sounds.
Note: Default setting is English
Battery Status
Battey Low
The Manikin battery is low.
Replace batteries soon.
OK
Replace Battey
The Manikin battery is VERY low.
Replace batteries immediately.
OK
Remote control battery status is indicated in the upper right corner
of the display. Batteries should be replaced immediately when batte-
ry empty
is indicated. When the batteries are totally empty,
the remote control will turn itself off without any warning. Replace
the remote control batteries with 4 AA size alkaline batteries.
Calibrate

Programming ECG

Set Running Rhythm
00:10 Logging [Log01]
A: Sinus
80
L.Lung: Normal
R.Lung: Normal
No Waiting Rhytm
Heart:
Defib. Enabled
Vocal:
BP: 120 / 80 P: N 4
Breathing Rate: 14
Extrasyst.
On the main display, use the left/right navigation buttons to highlight
the ECG functional area and press OK to edit the functions.
Some functions may be changed directly from the main display using
the quick execution buttons.
- HR+ and HR– change the heart rate up and down.
- Extrasyst. generates an extrasystole of the selected kind (on
Sinus rhythms only), immediately after the button is pushed. If no
extrasystole is selected, a Unifocal PVC will be generated.
Selecting Rhythm
Set Running Rhythm
1 Sinus rhythm
QRS
2 Ventr. Tachyardia
A
3 V.fib Very Coarse
4 V.fib Coarse
5 V.fib Medium
6 V.fib Fine
7 V.fib Very Fine
Set PEA
A
G
QRS
Using the navigation buttons, highlight the desired rhythm
and press OK.
Rate and QRS type can be modified using the quick execution
buttons,
OR
By highlighting the QRS or Rate area using the left/right navigation
buttons, and then use the numeric pad or up/down buttons.
QRS types can be set to A through G:
Available rates are dependent on the selected rhythm.
Setting PEA sets the BP to 0.
8
4
4
Normal
4
Normal
4
M
+
HR
Rate 80
+
HR