Herunterladen Inhalt Inhalt Diese Seite drucken

Hach 5500sc Wartung Und Fehlerbehebung Seite 7

Vorschau ausblenden Andere Handbücher für 5500sc:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
Clean the grab sample funnel
Clean the grab sample funnel before and after each use. Refer to
Figure 3 Clean the grab sample funnel
Replace the analyzer bottles
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
Replace the reagent(s) or standard(s) before the level in the analyzer bottle(s) is less than 10%.
1. Put the analyzer in shutdown mode. Refer to
2. When the status shows 100% completion, open the lower door.
3. Remove the cap from the reagent(s) or standard(s), then remove the bottle(s) from the analyzer.
4. Install the new analyzer bottle(s) and close the lower door. Refer to the operations manual.
5. Push menu and go to REAGENTS/STANDARDS.
6. Select RESET REAGENT LEVELS or RESET STANDARD LEVELS.
7. Select ENTER BLANK VALUE and enter the blank value from Reagent 1.
8. For reagents, select PRIME REAGENTS and confirm.
9. When the reagent prime is complete, start the analyzer. Refer to
operation
on page 4.
C A U T I O N
C A U T I O N
Put the analyzer in shutdown mode
Figure
3.
on page 4.
Put the analyzer back into
English 7

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis