8. Rectifying faults

Error messages can occur when
• the blood-pressure value is unusually high or low (
in the display),
• you move or talk during the measurement (
display),
• the cuff tube is not properly inserted (
play),
• inflation takes longer than 20 seconds (
display),
• the inflation pressure is higher than 300 mmHg (
in the display),
• an error occurs when storing the measured values (
appears in the display),
• exceeding measurement range (
In the about cases, you must repeat the measurement. Make sure
that the cuff tube is properly inserted and that you do not move or
talk. Put the batteries back in if necessary, or else replace them.

9. Specifications

Model no.
BM 16
Measurement
Oscillometric, non-invasive blood pressure
method
measurement on the upper arm
Measurement
Cuff pressure 0 – 299 mmHg,
range
systolic 40 – 250 mmHg,
diastolic 40 – 250 mmHg,
Pulse 40 –199 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
appears
appears in the
appears in the dis-
appears in the
appears
appears in the display).
Measurement
Max. permissible standard deviation accord-
inaccuracy
ing to clinical testing: systolic 8 mmHg /
diastolic 8 mmHg
Memory
2 x 50 memory spaces
Dimensions
L 156 mm x W 105 mm x H 74,5 mm
Weight
Approx. 315 g (without batteries)
Cuff size
24 to 36 cm
Permissible oper-
+10 °C to + 40 °C, 15 – 85 % relative air
ating conditions
humidity (non-condensing)
Permissible stor-
-10 °C to + 60 °C, 10 – 90 % relative air humid-
age and transport
ity, 700 –1060 hPa ambient pressure
conditions
Power supply
4 x 1,5 V
Battery life
For approx. 300 measurements, depend-
ing on the blood pressure level and/or pump
pressure
Accessories
Instruction for use, 4 x 1.5 V AA batteries,
storage pouch
Classification
Internal supply, IP21, no AP or APG, continu-
ous operation, type BF applied part
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to elec-
tromagnetic compatibility (EMC). Please note that portable
21
AA batteries