Press Process key.
As you are now in the [SET] menu item, press the Process key again.
Using the left/right keys, select the relevant digit and enter the value of the digit using
the Up/Down keys.
.0598518
Press the Operate key.
The display is now in "RUN: The filling volume is displayed".
To adjust the unit, use the Down key to change to "CAL: The current calibration con-
stant is displayed".
(CAL 1 / .0598518)
Press the Process key.
The memory location remains at 1.
Press the Process key.
Select using the [UNIT] menu item using the Down key.
Press the Process key.
Use the Up or Down key to set to "Kilogram".
Press the Operate key.
The display is now in "RUN: The filling volume is displayed".
To now switch to the calibration constant at memory location 0 (.0498765) with the Litre
unit, change the memory location now from 1 to 0.
Use the Down key to switch to "CAL: The current calibration constant is displayed".
Press Process key.
Use the Up or Down key to set the memory location to 0.
Press the Operate key.
The display is now in "RUN: The volume to be filled is displayed".
And back to kg with the calibration constant (.0598518)-->
Select memory location 1 again.
3. The user can delete the volume filled and also change the decimal place in the
display but cannot change the other settings.
To do so, enter settings in " SET: General Settings".
In chapter 5.2 "Change security level", the table shows that [LOC] Level 3 should be
set. All groups with the exception of RUN can only be processed by entering the PIN.
Use the Up key to switch to "SET: General Settings".
Press Process key.
As you are now in the [SET] menu item, press the Process key again.