4. CAL
Display of calibration constants
Setting range: 0.9999999 to 99.99999
4.1 Select the memory location of the calibration constants [CAL] 0 to 9.
There are 10 memory locations.
The settings of SET, UNIT and DEC can vary at each memory location. The displayed
unit of volume of the total volume display and the flow rate display depends on the set-
ting of the calibration constants at [UNIT]. This means that not only can the calibration
be changed by the selection of the memory location but also the volume unit.
4.2 Set the calibration constants at the selected memory location [SET].
4.3 Set the volume unit of the calibration constants at the selected memory location
[UNIT].
4.4 Set the decimal place of the calibration constants at the selected memory location
[DEC].
Process
4.1 Change mem-
4. CAL
Process
4.2 Change calibra-
tion constant [SET]
4.3 Change volume
4.4 Change decimal
The volume display and the total volume are deleted by changing the calibration
constants.
Select
ory location
Process
Process
unit [UNIT]
Process
place [DEC]
Attachment FLUXTRONIC FMJ 100
Operate and back to [RUN]
Select/Enter
7-digit
Operate and back
to [RUN]
Select
[-], kg, g, m³, mL, L,
lmp Gal, US Gal
Select
0 ... 3
31 / 60