Examples:
The volume meter is only suitable for use in normal mode.
The display is switched off when the FMC is delivered.
The display can be switched on again by pressing a key.
By pressing the Cancel key several times, the user can ensure that the unit has
changed over to "RUN: Filling volume is displayed".
1. The total volume should be displayed with a decimal place.
Use the Down key to switch to "TOT: The total volume is displayed"
Press Process key.
Select using the Down key [DEC].
Press Process key again.
Use the left key to set the flashing decimal point to one decimal place.
Press the Operate key.
The display is now in "RUN: The filling volume is displayed".
2. The volume unit is to be shown once in litres and another time in kilograms.
The volume unit is assigned to the calibration constant and is therefore set in "CAL: The
current calibration constant is displayed".
When delivered, a calibration constant is set at memory location 0, that was determined
in the factory using water (e.g. .0498765 L/pulse).
If the display is to show kg, either the density of the liquid must be known so that the
calibration constant can be converted from litre/pulse to kg/pulse, or weighing scales
are used to determine the mass of the volume filled.
For example the density of the liquid is 1.2 kg/L, therefore the new calibration constant
is 0.0498765L/pulse x 1.2 kg/L =0.0598518 kg/pulse.
Or
Use the volume meter to fill a sufficiently large volume into a bucket, for instance (dis-
play on the volume meter shows 25.4 litres). The weight of 30.48 kg can be read on the
scales. The density of the liquid is now known to be = 30.48 kg/ 25.4 litre =1.2 kg/litre.
Therefore again 0.0498765 L/pulse x 1.2 kg/L =0.0598518 kg/pulse
The new calibration constant is to be saved at memory location 1.
When the calibration constants are changed, the display of the volume filled and the to-
tal volume is reset.
Use the Down key to switch to "CAL: The current calibration constant is displayed".
Press Process key.
Use the Up or Down key to set the memory location to 1.
Attachment FLUXTRONIC FMJ 100
35 / 60