12. Technical data

Operating conditions:
10 - 40 °C / 50 - 104 °F
15% to 95% relative maximum humidity of air
Storage conditions:
(-20) – (+55) °C / (-4) – (+131) °F
15% to 95% relative maximum humidity of air
Weight: 115 g
Dimensions: 81 x 65 x 21 mm
Measuring method:
oscillometric, corresponding to the Korotkoff method:
phase I systolic, phase V diastolic
Measuring range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
34
Display range of cuff pressure: 0 – 299 mmHg
Measuring resolution: 1 mmHg
Statistic accuracy: pressure within ± 3 mmHg
Pulse accuracy: ± 5 % of the readout value
Voltage source: 2 x 1,5 V Alkaline-Batterien, Größe AAA
Battery life: approx. 320 measurements (new batteries)
IP class: IP20
Reference to standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Average useful life:
Monitor: 5 years or 10,000 measurements
Accessories: 2 years