Herunterladen Inhalt Inhalt Diese Seite drucken

Beurer TL 41 Gebrauchsanweisung Seite 66

Tageslichtlampe
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Immunity Test Standard
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, interruptions and
variations on power supply input
lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note: U
is the a.c. mians voltage prior to application of the test level.
T
IEC 60601-1-2 Test Level
± 8kV contact
± 2kV, ± 4kV, ± 8kV, ± 15kV air
± 2kV, for power supply lines
± 1kV signal input/ output
100 kHz repetition frequency
± 0.5kV, ± 1kV differential lines
± 0.5kV, ± 1kV, ± 1kV common mode
0 % U
; 0.5 cycle. At 0°, 45°, 90°,
T
135°, 180°, 225°, 270° and 315°.
0 % U
; 1 cycle and 70 % U
T
cycles; Single phase: at 0°.
0 % U
; 250/300 cycle
T
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
66
Compliance Level
± 8 kV contact
± 2kV, ± 4kV, ± 8kV, ± 15kV air
±2 kV, for power supply lines
Not applicable
100 kHz repetition frequency
± 0.5kV, ± 1kV differential lines
Not Applicable
0 % U
; 0.5 cycle. At 0°, 45°, 90°,
T
135°, 180°, 225°, 270° and 315°.
; 25/30
0 % U
; 1 cycle and 70 % U
T
T
25/30 cycles; Single phase: at 0°.
0 % U
; 250/300 cycle
T
30 A/m
50Hz/60Hz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
;
T

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis