4.3.2. DC Voltage measurement

1. Select the ""
2. Insert the test leads into the jack, the red plug into V/Ω jack, and black plug into COM
jack, like showed in Fig. 5.
3. Insert the two long ends of test leads to the desired circuit, then reading will be
displayed with automatic detection of the appropriate range.
4. If the reading is preceded by the "-" sign check this indicate that Voltage polarity its
reversed. Invert terminal leads for correct indication.
5. Press R-H key you can select manually the measurement ranges in cyclic order. Press
R-H key 2 seconds at least to come back to AUTOMATIC range selection.
O.L
6. The "
range. Press R-H key to select a higher range.
7. If the reading is difficult, press the D-H key to hold the obtained value. To exit from this
function press D-H key again.
wa
Quality HT Products Online at:
Maximum input for DC Voltage measurements is 600V.
Do not attempt to take any voltage measurement that exceeds the limits.
Exceeding the limits could cause electrical shock and damage the clamp
meter.
Fig. 5: Use of clamps for DC voltage measures
V
" position of selector functions.
" symbol means that the measured quantity is higher than the selected
As the clamp has high input impedance, it could happen that the meter
shows a continuous digits oscillation at display with missed signals on
inputs which is the result of the amplification of external noise. This is not a
defect, indeed the user can proceed with a test as the temporary displayed
value will not be added to the measured value.
www.
WARNING
WARNING
GlobalTestSupply
EN - 10
.com
sales@GlobalTestSupply.com
HT4012