Herunterladen Inhalt Inhalt Diese Seite drucken

struers TegraPol-21 Gebrauchsanweisung Seite 59

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Error message
TegraPol with TegraForce
TegraForce must be down for
preparation
TegraForce not locked
TegraForce must be up for manual
preparation
TegraForce must be up when
cleaning
Check water expeller on
TegraForce
TegraPol with TegraForce and TegraDoser-5
TegraDoser-5 nozzle block not
mounted
Method not valid anymore
Tube cleaning in progress
Suspension not applicable
Lubricant not applicable
TegraPol-21/ -25/ -31/ -35
TegraForce-3/ -5
Instruction Manual
Explanation
Nozzle block missing, or not
correctly seated.
Method has been edited on
TegraDoser-5.
The suspension specified in the
method has not been configured on
the TegraDoser-5.
The lubricant specified in the
method has not been configured on
the TegraDoser-5.
Action required
Press up/down button before
starting process.
Lock TegraForce with handle before
starting process.
Press up/down button before
starting manual process.
Press up/down button before
starting cleaning.
Press the release outlet valve to
drain the water/oil filter.
Make sure nozzle block is mounted
securely.
Re-read method by going back to
method list and select method
again.
Wait for cleaning process to finish.
Configure the suspension in one of
the pump positions, or select
another suspension in the method.
Configure the lubricant in one of the
pump positions, or select another
lubricant in the method.
53

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis