Error message
TegraPol
NVRAM initialised
Frequency inverter communication
error
Frequency inverter heat sink temp.
too high
Frequency inverter internal temp.
too high
Frequency inverter unknown error:
1
Frequency inverter type mismatch
error
Motor temperature too high
Motor temp. sensor disconnected
Cannot read keyboard
52
TegraPol-21/ -25/ -31/ -35
TegraForce-3/ -5
Instruction Manual

9. Trouble-shooting

Explanation
Default settings have been set for
all non-volatile parameters.
Internal frequency inverter error.
Fabrication error, incorrect motor
controller mounted in device.
Action required
None, message will be removed
after a few seconds.
Turn device off and on, and try
again. If problem persists call a
Struers service technician.
Turn device off and let it cool down
for a few minutes. Turn device on
and try again. If problem persists
call a Struers service technician.
Turn device off and let it cool down
for a few minutes. Turn device on
and try again. If problem persists
call a Struers service technician.
Turn device off and on, and try
again. If problem persists call a
Struers service technician.
Call a Struers service technician.
Stop process and let device cool
down for a few minutes. Try again. If
problem persists call a Struers
service technician.
Call a Struers service technician.
Call a Struers service technician.