Mounting and load application

3
Mounting and load application
The load cells are fixed at the mounting bores. For the recommended screws
and tightening torques refer to the table below:
Max. capacity
6...90 kg
1)
Recommended value for the stated property class. For screw dimensioning please refer to the
appropriate information given by the screw manufacturers.
Load must not be applied to the side where the cable connection is located, as
this would cause a force shunt.
Center of mounting plate
Recommended
flatness: 0.1
1: Mounting of PW22 at load introduction: 2 x M6; property class 10.9;
tightening torque 10 N·m; max. thread reach 20 mm.
Recommended flatness 0.1 mm on connection surface
2: Mounting of PW22 on base plate: 2 x M6; property class 10.9:
tightening torque 10 N·m; max. thread reach 20 mm
8
Thread
Min. property class
M6
A_6
Base plate
A2443-3.0 HBM: public
10.9
Loading direction
(1)
Mounting plate
PW22
(2)
Tightening torque
10 N@m
Load introduction
(Costumer construction)
1)
PW22

Werbung