In an emergency, the device func-
CAUTION
tions can be turned off by pressing
the "safe STOP" button (C) at the
front of the station. The LED bar (D) changed from green
color into red color.
A message appears in the display indicates that the station
was forced off. To restart, press the "ON/OFF" (A) or the
mains switch (2) off and on.
Adjust the safety temperature with screwdriver included
with the device.
Fig. 21
The station is controlled via a Wireless Controller (WiCo).
If the WiCo is mounted on the station, data is exchanged
between the station and WiCo via the contacts (G). The
screen of the WiCo displays the home icon .
The WiCo is equipped with a USB socket (Universal Serial
Bus) with which the WiCo can be connected with station,
the USB icon
appears on the screen.
If the WiCo is not connected with the station via a USB
cable, the data exchange between the station and the
WiCo via Bluetooth
. In this case, the Bluetooth
®
is displayed .

Setting the safety temperature

Useful information

icon
®
If the Bluetooth
®
function of the WiCois activated, the
Bluetooth
icon
appears on the screen and the user can
®
search the WiCo by pressing the searching key (E). Then,
a beep is heard.
The safety temperature setting will appear on the display.
Factory setting: maximum value
Adjustment range: 0 – 260 °C
DANGER
point of the fluid used.
Depending on the structure of the building, the Wireless
Controller can be operated at a distance up to 150 m from
the station, using the Bluetooth
The WiCo could be either installed on the station or put
on a safe place where is accessed easily by the user during
operation.
If the WiCo is mounted on the station, the battery is
charged through the contacts (G).
The battery could also be charged via the USB socket on
the Wireless Controller (See "Charging the Battery
Pack RB 1 (rechargeable battery)" in "Preparations"
section).
Note: The safe temperature
limit must always be set to at
least 25 ºC lower than the flash
connection.
®