Herunterladen Inhalt Inhalt Diese Seite drucken

Error Codes - IKA IC control Betriebsanleitung

Vorschau ausblenden Andere Handbücher für IC control:
Inhaltsverzeichnis

Werbung

Any malfunctions during operation will be identified by an error message on the display.
Proceed as follows in such cases:
 Switch off device using the main switch at the back of the device
 Carry out corrective measures
 Restart device
Error code
Effect
Error 01
Pump off
Heating off
Error 02
Pump off
Heating off
Error 03
Pump off
Heating off
Error 04
Pump off
Heating off
Error 05
Pump off
Heating off
Error 06
Pump off
Heating off
Error 07
Pump off
Heating off
Error 08
Pump off
Heating off
Error 09
Pump off
Heating off
Error 10
Pump off
Heating off
Error 11
Pump off
Heating off
Error 12
Pump off
Heating off
Error 13
Pump off
Heating off
Error 14
Pump off
Heating off
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.

Error codes

Cause
No External temperature sensor
Motor over current (rate current)
Motor over current (Max current)
Motor Hall signal missing
Too high liquid level
Too low liquid level
Too high voltage
Too low voltage
Device internal temperature is too
high
PC communication failure
Temperature difference between
control sensor and safety sensor is
too much
Safety temperature alarm
Heater switched off by safety circuit - Check safety temperature set value, liquid level
Fan error
Solution
Check this sensor
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
- Check the liquid level and buoyage
- Check the liquid level and buoyage
- Check the mains power
- Check the mains power
- Check the ambient temperature and let the
device cool down
- Check communication cable
- Check safety temperature circuit and bath fluid
- Check the bath temperature measurement
- Check the fan and clean the grids at the rear side
49

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hbc 5 controlHbc 10 control

Inhaltsverzeichnis