Herunterladen Inhalt Inhalt Diese Seite drucken

Setting Of The Safety Temperature; Maintenance And Cleaning; Associated Standards And Regulations - IKA EH 4 basic Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Setting of the safety temperature

The desired safety temperature can be adjusted with the „safe
temp" set screw. As soon as the desired safety temperature
is exceeded by a higher target temperature that has been set,
or if an error is encountered in the temperature controller, the
red pilot light (E) „Alarm" lights up. The heater turns off per-
manently in this state, and the pump continues to run.
The unit can only be placed in operation again by switching off
and then back on again wizh the rocker switch (A).
After the target temperature has been reached, the safety
temperature is determined as follows:
• Using an appropriate tool, turn the „safe Temp" set screw
(D) of the safety temperature controller until the switch-off
point to the left. The red pilot light (E) „Alarm" lights up.
• Then turn the „safe Temp" set screw (D) of the safety tem-
perature controller just slightly to the right.
• After you have turned the unit off and back on ag ain with
with the rocker switch , it is ready for operation.

Maintenance and cleaning

The IKA EH4 basic is maintenace-free. It is subject only to the
natural wear and tear of components and their statistical failu-
re rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and
the name of the spare part.
EH4b 1204
Please send in equipment for repair only after it has been cle-
aned and is free from any materials which may constitute a
health hazard. Use only cleansing agents which have been
approved by IKA to clean IKA devices. To remove use:
Dyes
isopropyl alcohol
Construction materials water containing tenside / isopropyl alcohol
Cosmetics
water containing tenside / isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information
from IKA. Wear the proper protective gloves during cleaning of
the devices.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Before using another than the recommended method for clea-
ning or decontamination, the user must ascertain with the
manufacturer that this method does not destroy the instru-
ment.
To avoid contamination, it is essential to change the bath liquid
from time to time.

Associated standards and regulations

Construction in accordance with the foolwing safety standards
DIN EN IEC 61010-1
UL 3101-1
Construction in accordance with the foolwing EMC standards
DIN EN IEC 61326-1
Associated EU guidlines
EMC-guidlines: 89/336/EG
Machine guidelines: 73/23/EG
CAN/CSA C22.2 (1010-1)