11. RUN a HbA1c TEST
STEP 6.
Insert the reagent pack in the cartridge and gently push into cartridge compartment
of the analyzer. The icon 'Close the lid' will be displayed.
NOTE: Do not force the reagent pack into the cartridge, it will only fit one
way.
STEP 7.
The test starts automatically when the lid is closed.
IMPORTANT: Do not open the lid during testing.
Do not move the analyzer during testing.
Do not expose the analyzer to vibration during testing.
STEP 8.
The HbA1c result displays after 5 minutes.
14