This breathalyzer is a high quality and sensitive measurement
device. To maintain the accuracy of the fuel cell sensor, it
should regularly be used. If the device is not used for an ex-
tended period of time, the sensor can dry out. ACE Instruments
recommends to use the breathalyzer at least 2 times per
month (even if you have not been drinking alcohol).
Due to changes in the product, the procedures stated can
change. This can take place without prior notice or adaptation
of the operating manual and is no cause for complaint, return
or price reduction. The user is responsible for recognizing
discrepancies between the operating manual and the actual
handling of the breathalyzer and must react accordingly.
Only the usage of original ACE Instruments mouthpieces
Guarantees test result accuracy
Guarantees 100 % hygenity
Guarantees full warranty of this device
ACE Instruments advises
even if you get a result meaning „safe range", you should
never drive after consuming alcohol or do any other activities
that are severely impaired by alcohol consumption and might
endanger your health and safety or the health and safety of
others.
Display indications
Display
Meaning
The battery is empty.
BAt
The testing cannot
be performed
because the battery
is low.
FLo
The sampling vol-
ume or blow power
is not enough to
analyze the breath
sample.
21
Action
Replace with new
batteries.
Replace with new
batteries.
Press the start button, the
countdown will start again,
and "BLOW" will appear
on the display. You can
now perform a new test.
Out
If no breath sample
was given within 3
minutes, the device
turns off automat-
ically.
Outc
The testing cannot
be performed
because the device
is turned on outside
of the operating
temperature.
Err2
"Err2" states that
fluid or saliva has
entered the device
and the sensor is
unable to evaluate it.
500
The total number
of breath tests per-
formed has reached
500. The device
should be calibrated
promptly to maintain
the warranty and
accuracy.
Calibration
To maintain the accuracy of the breathalyzer, ACE Instruments
recommends a calibration after 500 tests or approx. every
12 months (depending which incident happens first). For a
calibration service you can send the device to ACE Instruments
(see www.alkomat.net). The calibration is not mandatory, but
the accuracy of the breathalyzer will decrease steadily without
it and additionally the warranty for the sensor expires.
We recommend sending it as a parcel, so that in cases of
loss an inquiry or proof of delivery is possible.
Press power button to
restart the countdown
cycle and retest.
Repeat the test after
the unit has reached a
temperature within the
operation range (5°C up
to 40°C).
Wait 20 minutes and
then try again. If "Err2" is
displayed for a long time,
contact your seller for
customer service.
Contact your seller for a
calibration service. You
can send the device for
calibration to the service
address specified at
www.alkomat.net.