Error
n
- This error message appears if there is an
internal error in the GlassBuddy.
Solution: If the message Error is repeated,
the GlassBuddy should be
returned to the manufacturer for
inspection.
No Data
n
- This message appears if there is no data
in the memory that can be shown.
"Measurement in Basic mode not possible"
n
- This error message appears if types of glass
are being analysed which can only be
measured with GlassBuddy Plus.
Solution: Software Upgrade to
GlassBuddy Plus
11 Charge battery
When the battery is fully charged, the device can
work permanently for 8 hours. When charged, the
battery usually lasts for one to two weeks under
normal conditions of use. At a battery status of
approximately 10% you will be shown the message
Charge battery. Measurements can still be taken
but the GlassBuddy should be charged as soon as
possible to ensure problem-free operation.
Connect the GlassBuddy to your PC or laptop using
the supplied USB cable. To charge the battery at
the mains supply or in the car, we also have a USB
power set available as an accessory (BO 5164751);
this contains interchangeable plugs for a number of
countries (EU/US/UK/AU).
Using the car adapter you can charge the Glass-
Buddy whilst driving from one building site to the
next.
When the battery is almost empty, it takes approxi-
mately 120 minutes to charge it to 100 % again (if
the GlassBuddy remains switched on during char-
ging, the charging time is longer). The GlassBuddy
will indicate when the charging process is over. The
display will then show PC connection - battery is
charged. Nevertheless the GlassBuddy has to be
switched on for displaying this message.
The GlassBuddy contains a permanently installed
top quality Li-ion rechargeable battery with a very
long service life. If the performance of the battery
diminishes over the years, Bohle can replace the
battery.
12 Care and Maintenance
To avoid damage or malfunctions of your GlassBud-
dy, please strictly observe the following:
The GlassBuddy must always be stored in the
n
supplied safety case as this is the only way to
effectively prevent damage.
To avoid damage, the GlassBuddy must only
n
be operated and/or stored at temperatures
between 0° C and +40° C.
The GlassBuddy must not be exposed to
n
vibrations or impacts.
Only clean the display window and casing
n
with a dry, clean and soft cloth. Never use
cleaning liquids or thinning agents.
The laser opening and the sensor slit may only
be cleaned with a dry, clean and soft brush and/or
with clean compressed air.
Repairs may only be performed by the
n
manufacturer; please contact your dealer.