CONDITIONS OF USE

The B102 clamp must be used under the conditions defined above in order to meet the requirements
for operator safety and the metrological performance levels.
4.3.1
OVERLOADS
Ip limit current: permanent 400 AC RMS
Peak current: < 1000A.
Permissible transient di/dt: ≤30 A/µs.
Conductor temperature: ≤ 70° C with a maximum peak of 90° C.
4.3.2
FREQUENCY
For use from 48 Hz to 1 kHz.
4.3.3
ENVIRONMENTAL CONDITIONS
The graph shows the air temperature and humidity conditions for the casing.
: Reference range.
: Operating range.
: Storage range.
For use indoors.
Pollution level 2 to IEC 61010 .
Operating altitude: ≤2000m on uninsulated conductors.
Transport altitude: ≤12000m.