3 . H O W DO ES M I TECH NOLOGY W ORK
( M US C L E I NTEL L I GENC E)?
N.B.: The activation/deactivation of MI functions is performed via the Settings menu
MI-SCAN
Just before starting a workout session, the MI-scan function probes the chosen muscle group and
automatically adjusts the stimulator settings to the excitability of this area of the body, depending on your
physiology.
This function results in, at the start of the programme, a short test sequence during which measurements are
taken.
At the end of the test, intensities must be increased to start the programme.
MI-TENS
The MI-tens function limits unwanted muscle contractions in painful areas.
With each intensity increase, a test phase occurs and if a muscular contraction is detected, the device
automatically reduces the intensity of the stimulation.
This function is only accessible in the TENS, Epicondylitis and Tendinitis programmes.
MI-RANGE
The MI-range function allows the user to select the ideal level of stimulation intensity for recovery, massage,
capillarisation or even muscle pain programmes.
At the beginning of the programme, the device prompts the user to increase stimulation intensity. During this
increase, the device analyses the response of each stimulated muscle, and deduces their optimal level. As
soon as a muscle reaches its optimal level, the related channel is automatically deselected and the stimulation
intensity cannot be increased. To recover control, simply re-select the channel in question and raise or lower
the stimulation intensities.
MI-AUTORANGE
The MI-autorange function pursues the same objective as MI-range except that in this case everything is
done automatically.
At the beginning of the programme, a single press of the up key of the multifunction pad allows the device to
automatically increase the stimulation intensities until it detects the ideal level of intensity. Pressing the down
key of the multifunction pad cancels the MI-autorange function. The device then goes into manual mode and
intensities must be managed by the user.