Configuration

2.3
Approvals
The approvals are shown on the rating plate, see "Notes on Approval"
on page 348. The technical approvals are valid for the X-am 2500 gas
detection instrument and the calibration cradle. The explosion-
protection approvals are only valid for the X-am 2500 gas detection
instrument; the calibration cradle must not be used in the Ex zone.
The BVS 10 ATEX E 080 X technical suitability test is based on the
calibration with the target gas.
CE marking:
Electromagnetic compatibility
(Directive 2004/108/EC)
Explosion protection (Directive 94/9/EC)
3
Configuration
To individually configure an instrument with standard configuration,
the instrument must be connected to a PC via the USB infrared cable
(order no. 83 17 409). Dräger CC-Vision PC software is used to
perform the configuration. The Dräger CC Vision PC software can be
downloaded free of charge from the product page of the X-am 2500 at
the following Internet address: www.draeger.com.
Change configuration: see Technical Manual.
20
Standard instrument configuration:
®
1
Dräger X-am
2500
2
Bump test mode
2
Fresh air calibration
2 3
Operating signal
2
Switch off
2
LEL factor
(ch
)
4
2 4 5
STEL
(short-term average)
2 5 6
TWA
(shift average)
Can be acknowledged, non-latching, pre-alarm,
7
Alarm A1
Cannot be acknowledged, latching, like main
7
Alarm A1 at O
sensor
2
Cannot be acknowledged, latching, main alarm,
7
Alarm A2
®
1)
X-am
is a registered trademark of Dräger.
2)
Different settings can be selected to meet customer requirements on delivery.
The current setting can be checked and changed with the Dräger CC Vision software.
3)
A periodic short signal indicates the operating capacity of the instrument. If there is no
operating signal, correct operation cannot be guaranteed.
4)
STEL: average value of an exposure over a short period, generally 15 minutes.
5)
Interpretation only if the sensor is designed for this.
6)
TWA: shift averages are workplace limit values for generally eight hours per day of
exposure for five days a week during a working life.
7)
Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger
CC Vision PC software.
Extended bump test
ON
ON
allowed
4.4 (vol. %)
(4.4 vol. % corresponds to 100 %LEL)
STEL function – disabled
Average value duration = 15 minutes
TWA function – disabled
Average value duration = 8 hours
rising flank
alarm, falling flank
rising flank
Dräger X-am 2500