9.2 Securing the instrument

Under certain circumstances safe operation of the BENNING PV 1-1 is no longer ensured, for
example in the case of:
-
Visible damage of the casing.
-
Incorrect measurement results.
-
Recognisable consequences of prolonged storage under improper conditions.
-
Recognisable consequences of extraordinary transportation stress.
In such cases the BENNING PV 1-1 must be switched off immediately, disconnected from the
measuring points and secured to prevent further utilisation.
9.3 Cleaning
Clean the casing externally with a clean dry cloth (exception: special cleaning wipers). Avoid
using solvents and/ or scouring agents for cleaning the instrument. It is important to make sure
that the battery compartment and battery contacts are not contaminated by leaking electrolyte.
If electrolyte contamination or white deposits are present in the region of the batteries or battery
casing, clean them too with a dry cloth.

9.4 Battery replacement

Before opening the BENNING PV 1-1, make quite sure that it is voltage free!
c
Electrical danger!
The BENNING PV 1-1 is supplied by six 1.5 V batteries/ type AA (IEC LR6).
A battery change (see Figure 9) is required, if the battery symbol P appears in the display unit 1 .
Proceed as follows to replace the batteries:
-
Switch the BENNING PV 1-1 off.
-
Put the BENNING PV 1-1 face down and unscrew the screw of the battery compartment
cover.
-
Lift off the battery compartment cover (in the area of the housing slots) from the bottom part
of the battery compartment.
-
Remove the discharged batteries from the battery compartment.
-
Then, insert the batteries into the battery compartment at the provided places (please ob-
serve correct polarity of the batteries).
-
Lock the battery compartment cover into place on the bottom part and tighten the screw.
See figure 9:
Battery/ fuse replacement
Make your contribution to environmental protection! Do not dispose of
discharged batteries in the household garbage. Instead, take them to a
m
collecting point for discharged batteries and special waste material. Please
inform yourself in your community.
09/ 2012
etc.
or
BENNING PV 1-1
The electronic components of the BENNING PV 1-1
have reached the maximum admissible temperature.
Disconnect the BENNING PV 1-1 from the object to
be measured and let it cool down.
The DC short-circuit current exceeds the maximum
value of 10 A. Measurement has been stopped.
The DC open circuit voltage has exceeded the ma-
ximum value of 1000 V. The measurement has been
stopped.
The BENNING PV 1-1 is not correctly calibrated, see
chapter 9.6 "Calibration".
Please return the BENNING PV 1-1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
Please return the BENNING PV 1-1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
Please return the BENNING PV 1-1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
Please return the BENNING PV 1-1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.