4. Internal origin [SR-801R]

4-1. Organisation of origin position identification names

4-1-1. Effective length and origin position

1
2

4-1-2. Detection direction

The detection direction is categorized by the type of origin mark.
SR-801R
2500
Model
Effective length
Origin mark
The effective length and origin position are given on the effective length label.
Effective Length
The origin position is set with the following method.
q "R" stands for right and is the distance from the right effective length mark, a value in the
range from 0 to (effective length ÷ 2).
w "L" stands for left and is the distance from the left effective length mark, a value in the
range from 0 to {(effective length ÷ 2) - 1}.
e The central origin is only the Center, "C".
: Standard specification
: Reverse specification
R-200
(Standard)
Origin position
Detection direction
Effective length label
Absoiute
2500
Zero
Point
Effective length (mm)
R-200
Origin position (mm)
33