Herunterladen Inhalt Inhalt Diese Seite drucken

Tunze 7032/2 Gebrauchsanleitung Seite 55

Inhaltsverzeichnis

Werbung

Failure: Excessively high or low readings are
displayed.
Cause: The electrode is soiled or calibration is
too old.
Remedy: Clean the electrode (every one to
three months) and recalibrate (every three to
six months). The electrode should be mounted
in such a way that it is placed as dark as
possible.
Failure: Calibration impossible; reading does
not change.
Cause: Calibration solution is too old or soiled.
Remedy: Replace the calibration solution.
Cause: The electrode is soiled.
Remedy: Clean and recalibrate the electrode.
Cause: The electrode or the measuring
instrument is defective.
Remedy: Send the measuring instrument in for
repair.
Conductivity Meter 7032/2
Disfonctionnement : L'indicateur affiche des
valeurs trop hautes ou trop basses.
Raison : Electrode sale ou étalonnage trop
ancien.
Solution : Nettoyez (tous les 1 à 3 mois) et
étalonnez l'électrode (tous les 3 à 6 mois).
L'électrode doit être montée si possible en un
endroit sombre.
Disfonctionnement : Etalonnage impossible,
l'affichage varie dans une fourchette trop
étroite.
Raison : Solution étalon trop ancienne ou
souillée.
Solution : Remplacez la solution étalon.
Raison : Electrode sale.
Solution : Nettoyez et étalonnez l'électrode.
Raison : Electrode ou appareil de mesure
défectueux.
Solution : L'appareil de mesure nécessite un
envoi en réparation.


Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

7028/3

Inhaltsverzeichnis