Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration In Salt Water, Brackish Water - Tunze 7032/2 Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Calibration
If a salt water and a fresh water calibration is required,
the salt water calibration should be carried out first
as this will also change the fresh water calibration.

Calibration in salt water, brackish water

Switch the unit on (ON).
Use the „RANGE" button to select the 200 mS
scale (1).
Immerse the electrode into the 50 mS calibration
solution TUNZE® 7030.14 (2).
After 3 minutes use the screw-driver supplied long
with the unit to carefully turn the screw in the pper
opening in the right side wall (3) until the display (4)
renders the reading „50.0".
Hold the screw-driver vertically !
The unit is ready for use after the electrode has
been rinsed.
Etalonnage de l'électrode
Si un étalonnage eau douce et un étalonnage eau de
mer sont nécessaires, il est impératif d'entreprendre
tout d'abord celui pour l'eau de mer car il influencera
également l'étalonnage pour l'eau douce.
Etalonnage pour eau de
mer et eau saumâtre
Enclenchez l'appareil (ON).
A l'aide du bouton „RANGE", choisissez l'échelle
200 mS (1).
Plongez l'électrode dans la solution étalon 50 mS/
cm TUNZE® 7030.14 (2).
Après 3 minutes et à l'aide du tournevis fourni,
réglez la vis supérieure 50 mS (3) sur le côté droit de
l'appareil jusqu'à ce que l'afficheur indique 50,0 (4).
Le tournevis doit être tenu parfaitement droit !
Après rinçage de l'électrode, l'appareil est prêt à
l'emploi.