Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4. Plug the black and red test leads into the COM and V/Ω input
terminals respectively.
5. Touch the probes to the capacitor. Always observe polarity
makings when measuring pobrized capacitors.
6. Read capacitance value directly from the display.
7. After completing the measurement, disconnect the DCM test
leads.
6.7. Frequency measurements
1. (P 1610)
Set the rotary selector switch to the Hz position.
(P 1615)
Set the function switch to V/Hz%-position and
select Hz or %duty with the Hz%-button.
(P 1625)
Set the function switch to V/Hz-position and
press the MODE-button for >2 seconds to select
the frequency range.
2. Plug the black and red test leads into the COM and Hz input
terminals respectively.
3. Determine that the amplitude level of the signal to be measured
is not greater than the input voltage limit (250 V AC/DC). The
signal amplitude must also be greater than the sensitivity level.
4. Attach the probe tips to the points across which the frequency is
to be measured, and read the result directly from the display.
5. Disconnect the DCM test leads.
6.8. Temperature measurements
1. Set the function switch to ° C ° F or TEMP position. The meter
automatically defaults to ° C range.
2. Insert the meter´s white temperature adaptor into the V/Ω-input
terminal and the COM-terminal. Ensure that the minus marking at
the adaptor is inserted into the COM-terminal and the plus-
marking is inserted into the V/Ω-input terminal.
-45-

4. Spezifikationen

4.1. Gleichspannung
Modell
Bereich
Auflösung
P 1610
400 mV
100 µV
4
V
1 mV
40
V
10 mV
400
V
100 mV
600
V
1
P 1615
400 mV
100 µV
4
V
1 mV
40
V
10 mV
400
V
100 mV
600
V
1
P 1625
400 mV
100 µV
4
V
1 mV
40
V
10 mV
400
V
100 mV
600
V
1
Überlastschutz: 600V AC/DC
4.2. Wechselspannung
Modell
Bereich
Auflösung
P 1610
400 mV
100 µV
4
V
1 mV
40
V
10 mV
400
V
100 mV
600
V
1
P 1615
400 mV
100 µV
4
V
1 mV
40
V
10 mV
400
V
100 mV
600
V
1
P 1625
400 mV
100 µV
(Echt-
4
V
1 mV
Effektiv)
40
V
10 mV
400
V
100 mV
600
V
1
Überlastschutz:
600 V AC/DC
Frequenz-Bereich:
50/60 Hz
Genauigkeit
± 0,8% v.M. + 3 dgt
± 1,5% v.M. + 3 dgt
V
± 2,0% v.M. + 3 dgt
± 0,8% v.M. + 3 dgt
±1,5% v.M. + 3 dgt
V
± 2,0% v.M. + 3 dgt
± 0,8% v.M. + 2 dgt
± 1,5% v.M. + 2 dgt
V
± 2,0% v.M. + 2 dgt
Genauigkeit
± 0,8% v.M. + 20 dgt
± 1,8% v.M. + 5 dgt.
V
± 2,5% v.M. + 5 dgt.
± 0,8% v.M. + 20 dgt.
± 1,8% v.M. + 5 dgt.
V
± 2,5% v.M. + 5 dgt.
± 1,0% v.M. + 10 dgt.
± 1,5% v.M. + 8 dgt.
V
± 2,0% v.M. + 8 dgt.
-6-