pH measurement
(Pay attention to the technical data of the electrode additionally)
Measuring range
Resolution
Accuracy
General points
Usable electrodes
Calibration procedure
Calibration interval
Slope range
Offset voltage (Asymmetry)
Redox-Messung
Measuring ranges
Measuring range 1
Resolution
Accuracy
Measuring range 2
Resolution
Accuracy
Usable electrodes
Technical data
-2.00 ... 16.00
0.01
0.01 ± 1 digit
Input resistance: > 5 · 10
Input currence:
pH electrodes with plug according to DIN 19262
and electrode zero point pH = 7.0 ± 0.5
two- or three-point-calibration:
AutoCal TEC
AutoCal
one- or two-point-calibration:
ConCal
1 ... 999 days (preset 7 days)
-62 ... -50 mV/pH
-30 ... +30 mV
-199,9 mV ... + 199,9 mV
0,1 mV
0,2 mV ± 1 digit at operating temperature
15 °C ... 35 °C
-1250 mV ... + 1250 mV
1 mV
1 mV ± 1 digit
Pt-Electrodes with plug according to DIN 19262
S+
12
Ohm
5 · 10
-13
A typically
< 2 · 10
-12
A at 25°C
WTW Technical Buffer Solutions
Buffer Solutions according to
DIN 19 266
1 or 2 point calibration with
suitable buffer solutions
,21
85