• Unscrew the cap on the medication reser-
voir.

4.2 Controller

• Wipe the outer surfaces of the controller with a clean, damp cloth as necessary.
NOTICE
Liquids that get into the device can cause a fault in the device. Therefore, never
spray any liquids onto the controller or the power adapter.

4.3 Nebuliser

NOTICE
In order to avoid damaging the aerosol head, never attempt to process the medic-
ation reservoir in a microwave oven or a dishwasher. Mechanical cleaning by
brushing or scouring can also impair the function of the device.
Cleaning
The nebuliser, the mouthpiece and the mask must be cleaned thoroughly after every
application.
• Place all parts in warm tap water with a little
dishwashing liquid for at least 5 min.
• Rinse all parts thoroughly in running water.
• You can remove excess water more quickly
by shaking all parts.
Rinsing the aerosol head
The aerosol head is located in the medication reservoir. To ensure perfect functioning,
it should be rinsed once a week using the VELOXcare.
Rinsing the medication reservoir with the VELOXcare serves to mechanically
flush the membrane in the aerosol head. This rinsing does not replace the need
to clean and disinfect the medication reservoir!