Operation - RCD Testing
If the error number is again displayed, switch the
instrument to off, and return the instrument to the
manufacturer for service, together with a description of
the events leading to the message display. See
'Returning an Instrument for Repair'.
LCB2000/2 and LCB2500/2
Error 1
Number to be displayed is too large
Error 4
Event buffer overflow
Error 6
Rotary switch error, unstable output code
Error 7
Rotary switch error, invalid output code
Error 10
Zero cross error - fuse blown during test
Error 12
7109 a/d converter failed
Error 15
H8 a/d converter failed
Error 16
Loop test ranging error
Error 22
EEPROM failure, no acknowledge signal
Error 23
EEPROM data fault, serial number
checksum error
Error 24
EEPROM corrupted, end of data marker not
found
Error 33
PLD 'status' error
Error 34
False RCD trip
Error 35
EEPROM data fault, result checksum error
Error 48
EEPROM data fault, calibration constant
checksum error
Error 63
PLD 'status' timing error
Error 81
EEPROM data fault, error in stored foreign
language
Error 95
Low current loop test failed - PLD sequence
did not finish
LCB2500/2 only
Error 11
RS232 transmit data string too long
Error 17
RS232 transmit error
Error 31
RS232 receive error
27