Inserting the Module
Note: Be sure to connect the shielding properly!
The terminals 2 and 8
are covered by an ESD shield.
To connect the sensor cable, just
pull it back.
1.
Switch off power supply
2.
Open the device (loosen the 4 screws at the front)
3.
Place module in slot (D-SUB connector)
4.
Tighten fastening screws of the module
5.
Open ESD shielding cap (covering terminals 2 and 8)
6.
Connect sensor cable.
To avoid interferences, the cable shielding must be
completely covered by the ESD shielding cap.
7.
Close ESD shielding cap (covering terminals 2 and 8)
8.
Close device, tighten screws at the front
9.
Switch on power supply
10.
Set parameters
PH 3400(X)-035
Make sure that the cable
glands are tightly closed to
protect against humidity.